| Model Number |
MR8-AVA15-R |
| Medical Device Problem Codes |
Corroded (1131); Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis: evaluation determined that the an overheating test could not be conducted due to another device malfunction.The likely cause was identified as distal bearings are stuck/sticking.It was also noted spacers are corroded.B3: date of notification: (b)(6) 2024.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Repair request initiated for device with the report of overheating and corrosion.It was reported that there was no patient or staff impact.Repair is being escalated to product event due to reason for the return.
|
| |
|
Search Alerts/Recalls
|