(b)(4).D4: product id was provided for three screws however, it is unknown which of the three screws has migrated.Therefore, the migrated screw could be any of the following blunt tip screw, ã¿ 4x46mm item# 47248604640 lot# 3155779, udi: (b)(4), manufacturing date: apr 11, 2023, expiration date: apr 11, 2028.Blunt tip screw, ã¿ 4x46mm item# 47248604640 lot# 3155809, udi: (b)(4), manufacturing date: apr 19, 2023, expiration date: apr 19, 2028.Blunt tip screw, ã¿ 4x48mm item# 47248604840 lot# 3185174, udi: (b)(4), manufacturing date: nov 30, 2023, expiration date: nov 30, 2028.D10: proximal humerus, right, long, ã¿ 10x220mm item# 47249622009 lot# 3136330, blunt tip screw, ã¿ 4x46mm item# 47248604640 lot# 3155809, blunt tip screw, ã¿ 4x48mm item# 47248604840 lot# 3185174, ann cort bone screw 4 x 28mm item# 47248612840 lot# 3185202, ann cort bone screw 4 x 28mm item# 47248612840 lot# 3191519, affixus ph nl cap 0mm item# 47248801000 lot# 3193703.G2: foreign: japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental mdr will be submitted.
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, g3, g6, h2, h3, h6, h10, h11.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Review of complaint histories identified additional similar complaints for the reported products and additional similar complaints for the reported part and lot combinations.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.It can be assumed that multiple contributing factors, related to either patient condition and behavior or implantation procedure, contributed to the migration of the screws.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.An additional deeper investigation was performed which identified the design limitation of the corelock mechanism of the affixus nail as a potential contributing factor.As part of the deeper investigation, a scientific literature search was conducted and a systematic review of the outcome of intramedullary nailing for acute proximal humerus fracture showed that screw migration and perforation into the joint is the second most common complication following secondary loss of reduction.In addition, scientific literature of competitor and predicate devices were assessed.This review has shown that the affixus® natural nail® proximal humeral system performs better and/or in equal range in comparison with legally marketed similar devices.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.No corrective actions, preventive actions, or field actions resulted after investigation of this event as this issue was already addressed in a corrective action.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|