EXACTECH, INC. NV GXL LNR, LIPPED 32MM ID GROUP 2 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 132-32-52
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/19/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10 concomitant devices (b)(6) 38h -flex drill m 3.2x38 hd lenkbar ((b)(6) 132-32-52 - nv gxl lnr, lipped 32mm id group 2 cups (b)(6) 180-65-25 - alteon 6.5mm screw, 25mm (b)(6) 142-32-93 - cocr fem head 32mm -3.5 offset 12/14 (b)(6) 188-00-05 - wedge plasma s/o sz 5 (b)(6) 186-01-52 - integrip cc, cluster 52mm, g2.The product associated with the reported event is within the scope of recall z-1729-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported via legal documentation that approximately 41 months after a right total hip replacement procedure, the patient underwent a revision procedure to address prosthesis wear.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The most likely cause for the reported revision due to prosthesis wear is a combination of the risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).Additionally, the sequence of events as related to the reported wear cannot be confirmed and the contributions of each to the reported event cannot be determined.However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LNR, LIPPED 32MM ID GROUP 2 CUPS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville 32653
3523771140
MDR Report Key20179854
Report Number1038671-2024-03407
Device Sequence Number8931765
Product Code JDI
UDI-Device Identifier10885862023315
UDI-Public10885862023315
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K070479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/14/2020
Device Catalogue Number132-32-52
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2024
Supplement Date Received by Manufacturer02/11/2025
Initial Report FDA Received Date09/10/2024
Supplement Report FDA Received Date02/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/17/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-