MAUDE Adverse Event Report: SMITH & NEPHEW, INC. KIT, STABILIZATION SHOULDER, (12); ACCESSORIES, OPERATING-ROOM, TABLE (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. KIT, STABILIZATION SHOULDER, (12); ACCESSORIES, OPERATING-ROOM, TABLE (KIT) Back to Search Results
Catalog Number 7210573
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/11/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference (b)(4).H10, h3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that during set up, after the user opened the shoulder stabilization kit, it was found that it was dirty, it had finger marks (similar to fingerprints).The finger prints were found in the internal package.The material was not used in the surgery.There was a smith and nephew back-up device available and no surgical delay was reported.There was no patient involvement.
 
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).A device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the sterile packaging process found that the assembler should visually inspect the tyvek pouch and the sterilization wrap for particulates.Please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.If any additional information related to this event is received in the future, this investigation will be reopened for evaluation.
 
Event or Problem Description
It was reported that during set up, after the user opened the shoulder stabilization kit, it was found that it was dirty, it had finger marks (similar to fingerprints).The finger prints were found in the internal package, in the nearly wrap.The material was not used in the surgery.There was a smith and nephew back-up device available and no surgical delay was reported.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT, STABILIZATION SHOULDER, (12)
Common Device Name
ACCESSORIES, OPERATING-ROOM, TABLE (KIT)
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key20181887
Report Number3003604053-2024-00051
Device Sequence Number8696084
Product Code FWZ
UDI-Device Identifier03596010511232
UDI-Public03596010511232
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Catalogue Number7210573
Device Lot Number4073LR0400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2024
Supplement Date Received by Manufacturer09/13/2024
Initial Report FDA Received Date09/10/2024
Supplement Report FDA Received Date09/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-