• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH VIPER PRIME CFXFN XTB 6X50MM S; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH VIPER PRIME CFXFN XTB 6X50MM S; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD Back to Search Results
Catalog Number 186760450S
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the screw could no longer be released from the viper prime screwdriver.There was no surgical delay and no medical intervention required.The procedure was successfully completed.This report is for one (1) viper prime cfxfn xtb 6x50mm s.This is report 1 of 2 for complaint (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9, d4 (expiry date), d4 (lot), h4 corrected: d4 (primary udi number) h3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6: the product was returned to j&j medtech orthopaedics for evaluation.The j&j medtech orthopaedics team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the screwdriver and screw were observed that the devices were returned coupled with significant wear marks.In the functional test, a higher-than-normal force was required for disassembly; however, the devices were successfully separated without issue.Upon inspection, no damage was found on the threads where the devices are fastened, and therefore, the alleged issue could not be confirmed.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was unconfirmed as the observed condition of the viper prime cfxfn xtb 6x50mm s would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Device history review (dhr): a manufacturing record evaluation was performed for the finished device product code: 186760450s lot number: tbanhw it was electronically reviewed and no nonconformances/manufacturing irregularities were identified during the manufacturing process.The product was released on: 11-apr-2024 manufacturing site: jabil le locle expiry date:28-feb-2029 if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER PRIME CFXFN XTB 6X50MM S
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle 02400
SZ   02400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
8472871282
MDR Report Key20182576
Report Number1526439-2024-02350
Device Sequence Number15519152
Product Code NKB
UDI-Device Identifier10705034530109
UDI-Public(01)10705034530109(17)290228(10)TBANHW
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K173095
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number186760450S
Device Lot NumberTBANHW
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/17/2024
Supplement Date Received by Manufacturer09/18/2024
10/16/2024
Initial Report FDA Received Date09/10/2024
Supplement Report FDA Received Date09/26/2024
10/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-