| Model Number |
VTICM5_12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
04/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Manufacturer narrative: refractive surprise: each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that an implantable collamer lens was implanted into the patient's eye.The patient experienced refractive surprise.The lens was exchanged with a longer length, different axis lens and the problem was resolved.The cause of the event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: returned product evaluation: lens was returned in a microcentrifuge vial, dry with residue/debris on the product.Visual inspection found the lens haptic bent and fiber on the lens.Dimensional and functional inspection found the lens within specifications.Claim#: (b)(4).
|
| |
|