| Model Number |
PCF-H190I |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed during the device inspection, that the colonovideoscope exhibited foreign material inside jet tube.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Based on the results of the investigation, the definitive root cause of the presence of foreign material could not be determined.Additional information about the event was requested but not received.It is unknown if the user deviated from the reprocessing instruction manual.The instruction manual identifies detective and preventative verbiage associated with foreign material in ¿gif/cf/pcf-190 series operation manual chapter 3 preparation and inspection¿ and ¿gif/cf/pcf-190 series reprocessing manual chapter 5 reprocessing the endoscope.¿ olympus will continue to monitor field performance for this device.
|
| |
|