• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA SURGICAL OPTICS, INC. B1PC; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA SURGICAL OPTICS, INC. B1PC; INTRAOCULAR LENS Back to Search Results
Model Number B1PC (+ 22 D)
Medical Device Problem Code Sharp Edges (4013)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Date of Event 08/05/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This initial emdr is being submitted to fda for an event that occurred in the usa.Haptic damage is indicated as a potential malfunction related to the iol, as stated under the warnings section of the product's instructions for use (ifu).Regarding section h6 - manufacturer's codes for: type of investigation, findings, and conclusion are pending device return and completion of product investigation.Once the product investigation is completed, a follow-up report will be submitted to fda which will include the manufacturer's codes for type of investigation, findings, and conclusion.
 
Event or Problem Description
Event occurred in usa.Damaged haptic after implantation.When he went to spin the lens into place it appeared the haptics tore the capsule.Root; broken haptic during use; haptic does not fold as per ifu.Capsular bag tear / rupture.Product remained in eye after clinical assessment.Patient health not impacted.Patient has recovered and is fine.He left lens implanted.But questioned whether the haptics had changed.Lens deployed as normal.When he went to rotate the haptics tore the bag.Problem code: a051102 - sharp edges.
 
Additional Manufacturer Narrative
This follow-up report #1 emdr is being submitted to fda for a reportable event that occurred inside the usa.The report includes additional information not available/included in the initial report.Additional information: g6: type of report - noted as follow-up #1.H2: type of follow-up - noted for additional information.H6: added codes for manufacturer's investigation: type; findings; and conclusion.The product was not returned to the manufacturer.The investigation was conducted, with the methods and results as noted below.No abnormalities were found in production and inspection records of the product.(serial no.: (b)(6); model: b1pc).We also confirmed there were no abnormalities on the loop pull strength test record of the material lot: (sowf-g115-10).From our investigation, we couldn't confirm the reported event.The exact root cause was not determined.However, based on the available information and our investigation, we believe this event was not caused by our product quality.A review of the most recent complaint trending data indicates that no significant trends have been identified at this time and no capa is required as part of the product evaluation.
 
Event or Problem Description
Event occurred in usa.Damaged haptic after implantation.When he went to spin the lens into place it appeared the haptics tore the capsule.Root; broken haptic during use; haptic does not fold as per ifu.Capsular bag tear / rupture.Product remained in eye after clinical assessment.Patient health not impacted.Patient has recovered and is fine.He left lens implanted.But questioned whether the haptics had changed.Lens deployed as normal.When he went to rotate the haptics tore the bag.Problem code: a051102 - sharp edges.
 
Additional Manufacturer Narrative
This follow-up report #2 emdr is being submitted to fda for a reportable event that occurred inside the usa.The report includes additional information not available/included in follow-up report #1.Additional information: h6: added codes for manufacturer's investigation: type; findings; and conclusion.
 
Event or Problem Description
Event occurred in usa.Damaged haptic after implantation.When he went to spin the lens into place it appeared the haptics tore the capsule.Root; broken haptic during use; haptic does not fold as per ifu.Capsular bag tear / rupture.Product remained in eye after clinical assessment.Patient health not impacted.Patient has recovered and is fine.He left lens implanted.But questioned whether the haptics had changed.Lens deployed as normal.When he went to rotate the haptics tore the bag.Problem code: a051102 - sharp edges.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
B1PC
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
HOYA SURGICAL OPTICS, INC.
110 progress
suite 175
irvine CA 92618
Manufacturer (Section G)
HOYA MEDICAL SINGAPORE PTE LTD
455a jalan ahmad ibrahim
singapore, singapore 63993 9
SN   639939
Manufacturer Contact
heber enriquez
110 progress
suite 175
irvine, CA 92618
MDR Report Key20185361
Report Number3006723646-2024-00098
Device Sequence Number8722667
Product Code HQL
UDI-Device Identifier00886158017240
UDI-Public00886158017240
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P080004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model NumberB1PC (+ 22 D)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2024
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date09/10/2024
Supplement Report FDA Received Date10/22/2024
10/22/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/31/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-