Zimvie did not receive one (1) tsx6b11, (tsx¿ implant, 6.0mmd, 11.5mml) for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Functional testing to recreate the reported event could not be performed due to the nature of the device & event.This complaint refers to the specific device being investigated for this complaint record.Dhr review was completed for the subject lot number 1279704.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1279704 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿packaging other¿ the customer did not submit images for the reported event.Based on the investigation and risk management file review as per rmf rm-00308-pfmea rev.1, the most likely root cause determined from the investigation is improper handling of devices/packaging outside of zimvie control.Therefore, based on the available information, a packaging malfunction could not be verified.Without device receipt, the reported event/coob was non-verifiable since the condition of the product/packaging received by the customer was unknown / non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report g3: date received by manufacturer g6: checked "follow-up" h3: changed "yes" to "no" h6: entered evaluation codes h10: added manufacturer narrative h3 other text : device was not returned.
|