| Model Number |
BML-V442QR-30 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported, that during a therapeutic lithotripsy.The mechanical lithotriptor, tip that was used to crush the stone broke.The procedure was completed on a similar device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction b5** this supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of this event was that the guide wire's distal end did not meet strength specifications due to suboptimal molding conditions during manufacturing.However, a definitive root cause could not be determined.The event can be prevented by following the instructions for use which state: the instruction manual contains a warning to the user regarding this event. (drawing number:gk5911, revision number : 23) ·when using the guide wire, insert the instrument with its distal tip in parallel with the guide wire while holding the distal tip as shown in figure 4.20.Be careful not to forcibly insert the instrument with a sharp angle between the distal tip and the guide wire as shown in figure 4.21.This may damage the distal tip.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was reported the mechanical lithotriptor's tip that was used to crush the stone, broke off.The issue occurred during a therapeutic lithotripsy procedure and was completed with a similar device.There were no reports of patient harm.
|
| |
|