Catalog Number 347-08-703
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Implant Pain (4561)
Date of Event 08/13/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - due to patient complained of left knee pain.
 
Additional Manufacturer Narrative
The agent reported "(patient complained of left knee pain s/p uni knee)".The previous surgery and the surgery detailed in this event occurred 3 years and 2 months apart.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.First, there was an ncmr#52277 associated with the concomitant part # 247-02-103, empowr partial kneetm, femur, np, size3, rm-ll which document that out of 9 parts lot, all parts were rejected due to location and orientation of laser etchings on part are incorrect.Later, the rejected parts were accepted to use as is after proper justification.Second, there was an ncmr#53372 associated with the same part which document that out of 9 parts lot, all parts were rejected due to routing and bom is incorrect.Later, the rejected parts were reworked and accepted after proper justification.Third, there was an ncmr#54185 associated with the same part which document that out of 9 parts lot, all parts were rejected due to fixturing method for overlay is inadequate.Later, the rejected parts were reworked and accepted after proper justification.Fourth, there was an ncmr#54775 associated with the same part which document that out of 9 parts lot, lot 2 parts were rejected due to out of tolerance.Later, the rejected parts were reworked and accepted after proper justification.All other items in the respective lot were met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.The root cause of this complaint was a revision surgery due to pain.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities or trauma.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
EMPOWR PARTIAL KNEETM, TIBIAL INSERT, SIZE 3, 8MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd
austin, TX 78758
MDR Report Key20186462
Report Number1644408-2024-01391
Device Sequence Number8032063
Product Code HSX
UDI-Device Identifier00190446294887
UDI-Public00190446294887
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K191325
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/20/2025
Device Catalogue Number347-08-703
Device Lot Number165Z1016
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2024
Supplement Date Received by Manufacturer10/03/2024
Initial Report FDA Received Date09/10/2024
Supplement Report FDA Received Date11/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/22/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
247-02-103 LOT: 159Z1003.; 357-02-103 LOT: 195Z1003.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR