| Model Number |
CR2 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer contacted stryker to report that their device was missing its lid magnet.In this state, the device would not detect the lid being opened or closed, and therefore would not automatically power on or off which can lead to a use error resulting in a failure to deliver defibrillation.There was no report of patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's device was not returned to stryker for evaluation.The customer will receive replacement lid and battery.A cause of the reported issue could not be determined.
|
| |
|
Search Alerts/Recalls
|