• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I TSV BELLATEK ENCODE EMERGENCE HEALING ABUTMENT 3.5MM(D) 3.8MM(P) 5MM(H); DENTAL ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I TSV BELLATEK ENCODE EMERGENCE HEALING ABUTMENT 3.5MM(D) 3.8MM(P) 5MM(H); DENTAL ABUTMENT Back to Search Results
Catalog Number TEEHA335
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: weight unknown / not provided product has been received by zimvie and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient came to see the surgeon and when the surgeon went to remove the encode abutment, the implant came out with it.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111 and 3331.H6: investigation findings code was added: 180.H6: investigation conclusions code was added: 4307.H10: narrative/data was updated.Zimvie received the reported healing abutment for evaluation.Visual evaluation was performed, implant has signs of use with bone debris on its external threads.Unable to disengage abutment from implant.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was customer error.Therefore, based on the available information, device malfunction did occur.The abutment was stuck to the implant and couldn¿t be released.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TSV BELLATEK ENCODE EMERGENCE HEALING ABUTMENT 3.5MM(D) 3.8MM(P) 5MM(H)
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key20189246
Report Number0001038806-2024-02027
Device Sequence Number8138493
Product Code NHA
UDI-Device Identifier00844868045731
UDI-Public(01)00844868045731(17)290530(10)1282140
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K212730
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 01/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberTEEHA335
Device Lot Number1282140
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/05/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/11/2024
Supplement Report FDA Received Date01/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TSX37B11, TSX IMPLANT, 3.7MMD, 11.5MML
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-