| Model Number |
15-2814/02 |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Notified by t.A.That on (b)(6) 2024, dr.D.Was performing a revision using the endo w hinge system and 2 lots of tibial augments were opened and each had a screw that would not disengage.Cement was then used to set the augment in place.On part 15-2814/02 lot 230619/2153, only the lateral screw could be removed.On part 15-2814/02 lot 230703/1306 , only the medial screw could be removed.It was also noted that the screws that were removed were very difficult to get out, and a 2.5 screwdriver was broken in the process.Lot 230703/1306 was cemented into place.Waiting information on return of lot 230619/2153.Correction 2024-08-21:initial pn / lot number information received was inaccurate.I have received the complaint sample and it is lot 230703/1306, so lot 230619/2153 is the part that has been implanted, which was verified by the representative.Complaint sample will be shipped for investigation.[customer].
|
| |
|
Additional Manufacturer Narrative
|
|
This is the final supplemental report.The complaint is closed.
|
| |
|
Event or Problem Description
|
|
Notified by (b)(6).That on (b)(6) 2024, dr.(b)(6).Was performing a revision using the endo w hinge system and 2 lots of tibial augments were opened and each had a screw that would not disengage.Cement was then used to set the augment in place.On part 15-2814/02 lot 230619/2153, only the lateral screw could be removed.On part 15-2814/02 lot 230703/1306, only the medial screw could be removed.It was also noted that the screws that were removed were very difficult to get out, and a 2.5 screwdriver was broken in the process.Lot 230703/1306 was cemented into place.Waiting information on return of lot 230619/2153.Correction 2024-08-21: initial pn / lot number information received was inaccurate.I have received the complaint sample and it is lot 230703/1306, so lot 230619/2153 is the part that has been implanted, which was verified by the representative.Complaint sample will be shipped for investigation.[customer].
|
| |
|
Search Alerts/Recalls
|