| Catalog Number |
SAC-00820 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "during our labeling process it was discovered seven units that were bent." there was no patient involvement.Associated mdr numbers include: 3006425876-2024-00884, 3006425876-2024-00898, 3006425876-2024-00897, 3006425876-2024-00896, 3006425876-2024-00895, 3006425876-2024-00893.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer provided 15 photos for analysis.The report of a kinked guide wire was confirmed through the photos as it revealed kinking/bending of the guide wire body and creasing of the packaging.The customer also returned one unopened sac kit for analysis.No evidence of use was observed.The kit was opened to further analyze the components.Visual inspection revealed no obvious kinking of the guide wire.Several creases on the product packaging and bending of the swg protective tubing were observed.The damage observed is consistent with defects related to storage and shipping.Both welds were present and appeared full and spherical.The packaging additionally clearly states, "do not bend".Functional inspection of the guide wire was performed per the instructions-for-use (ifu) provided with the kit which states , "insert desired tip of spring-wire guide through introducer needle into artery (until depth-marking [if provided] on wire enters hub of needle).Thread tip of indwelling catheter over spring-wire guide." the undamaged portions of the guide wire were threaded through the returned 20 ga introducer needle, and little to no resistance was experienced.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "precaution: use of excessive force may damage catheter or guidewire.Do not use if package is damaged." the customer report of a kinked guide wire prior to use was confirmed through complaint investigation of the returned sample.Several folds and creases were observed on the outside packaging and swg protective tubing.The guide wire met all relevant functional requirements, and a device history record review was performed with no relevant findings.The lidstock clearly states "do not bend." based on the customer description and the sample received, storage and shipping caused or contributed to this event.Teleflex will continue to monitor and trend for complaints of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that "during our labeling process it was discovered seven units that were bent." there was no patient involvement.Associated mdr numbers include: 3006425876-2024-00884, 3006425876-2024-00898, 3006425876-2024-00897, 3006425876-2024-00896, 3006425876-2024-00895, 3006425876-2024-00894, and 3006425876-2024-00893.
|
| |
|
Search Alerts/Recalls
|