| Model Number |
2900162 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from user facility via manufacturing representative regarding a product identified during an event.It was reported that curette was damaged during set assembly.No patient was involved.No further complications were reported/anticipated.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis #808722467; product id#2900162; lot# im04g015 visual and optical examination the ring at the tip of the curette has been broken into.This is consistent with over-torque e1:initial reporter-unknown medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|