COOK INC RETRACTA DETACHABLE EMBOLIZATION COIL; KRD DEVICE, EMBOLIZATION, VASCULAR Back to Search Results
Model Number G34199
Medical Device Problem Code Unraveled Material (1664)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1 - customer (person): postal code: k9j 7c6.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported that the delivery wire of a retracta detachable embolization coil unraveled.The device was required for a left internal iliac artery aneurysm embolization prior to graft placement.During the procedure, the delivery wire was advanced partially into the catheter when the hub of delivery cannula was inadvertently pulled on, separated, and could not be reattached.The delivery wire was then retracted toward the delivery cannula when a 5 cm portion began to elongate and unravel.The device was removed from the patient and discarded.A second retracta coil was opened and experienced delivery cannula hub separation.Finally, a third coil was obtained and was successfully deployed.A representative was present for the procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged or unavailable.Investigation ¿ evaluation it was reported that the delivery wire of a retracta detachable embolization coil unraveled.The device was required for a left internal iliac artery aneurysm embolization prior to graft placement.During the procedure, the delivery wire was advanced partially into the catheter when the hub of delivery cannula was inadvertently pulled on, separated, and could not be reattached.The delivery wire was then retracted toward the delivery cannula when a 5 cm portion began to elongate and unravel.The device was removed from the patient and discarded.A second retracta coil was opened and experienced delivery cannula hub separation.Finally, a third coil was obtained and was successfully deployed.A representative was present for the procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Reviews of the documentation, including the complaint history, device history record, quality control procedures, and instructions for use (ifu) for device, were conducted during the investigation.The complaint device was not returned for evaluation; therefore, a physical examination could not be conducted.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) for the device found no relevant nonconformances that could have contributed to the reported failure mode.It should be noted that there as one other complaint associated with the final product lot number from the same customer for a different failure on the same date.Cook also reviewed product labeling: the current instructions for use [t_mwcer_rev6] state the following: "under fluoroscopic visualization, slowly advance the delivery wire until the entire length of coil exits the distal end of the catheter.Ensure that the junction remains positioned just inside the catheter tip.Note: advancing the delivery wire slowly allows the junction to be seen more easily and reduces the risk of damaging it.Note: if significant resistance is encountered during coil advancement, do not continue advancing.Retract the delivery wire slightly, then gently re-advance it.If there is still significant resistance, withdraw the delivery wire from the catheter and try using a new coil with a shorter length." evidence gathered upon review of the dmr, ifu, and the dhr suggests that the device was not manufactured out of specification, and that there are no nonconforming devices in house or out in the field.Based on the information provided, no product returned, and the results of our investigation, it was concluded that a component failure unrelated to manufacturing or design deficiencies contributed to the reported event.It is possible that manipulation of the device contributed to the wire unraveling, but this could not be definitively confirmed.Per the risk assessment no further action is required.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETRACTA DETACHABLE EMBOLIZATION COIL
Common Device Name
KRD DEVICE, EMBOLIZATION, VASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key20190531
Report Number1820334-2024-01207
Device Sequence Number8138053
Product Code KRD
UDI-Device Identifier10827002341996
UDI-Public(01)10827002341996(17)290701(10)16136678
Combination Product (Y/N)N
PMA/510(K) Number
K151676
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG34199
Device Catalogue NumberMWCER-35-14-10
Device Lot Number16136678
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/28/2024
Supplement Date Received by Manufacturer03/13/2025
Initial Report FDA Received Date09/11/2024
Supplement Report FDA Received Date03/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-