Product complaint # (b)(4).Date sent to the fda: 9/11/2024.H6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: can you identify the lot number of the product that was used? tkbdea.¿ please provide the source or name and title of the external person providing answers to follow-up (external person submitting answers to sales rep) joy walters - lead nurse - also met with and spoke to sa¿s in the room.¿ please provide the status of the device(s) as it has not been received for analysis.If the device has been shipped, please provide the shipment tracking details.Part 1 of shipper kit has been returned.Second part shipper kit was delivered today.Will ship out tomorrow.H3 investigational summary: the product was returned to ethicon for evaluation.Four unopened samples were received for analysis.In order to evaluate the condition of the returned samples, the packets were opened.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during the evaluation.The functional test was performed using instron equipment and the pull force result was above the minimum requirements.
|