• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD MAXZERO MULTI FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD MAXZERO MULTI FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number MZXT9001
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd maxzero multi fuse extension set with needleless connector was damaged.The following information was received by the initial reporter with the following verbatim: it was reported by customer that t piece connector found broken.
 
Additional Manufacturer Narrative
The customer reported t-connector fell off, and returned one sample of material mzxt9001, lot 24039349.Upon initial visual examination, the set had a separated luer collar and the complaint is verified.There was no damage to the t-connector, just a small lip on the plastic of the luer post.The lip is designed to prevent the collar from falling of the luer post, but the dimension is too small, and is out of tolerance.The manufacturing site confirmed the root cause for loose collar is the dimension out of tolerance, and traced that cause to molding cavity #2.A containment action was implemented on 04sep2024 to contain lots in process from molding and tolerance manufacturing.A work order was issued 04oct2024 to replace the slide core pin 03 on the molding cavity #2.A quality alert was issued 15oct2024 to communicate and reinforce the correct molding procedure of the male luer component.Device history record review for model mzxt9001 lot number 24039349 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 22mar2024.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
 
Event or Problem Description
No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXZERO MULTI FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key20191310
Report Number9616066-2024-01284
Device Sequence Number7058354
Product Code FPA
UDI-Device Identifier10885403240805
UDI-Public(01)10885403240805
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K140831
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberMZXT9001
Device Lot Number24039349
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2024
Supplement Date Received by Manufacturer11/15/2024
Initial Report FDA Received Date09/11/2024
Supplement Report FDA Received Date11/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-