| Model Number |
AVA10 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Repair request initiated for device with the report of overheating.No patient impact reported.Repair is being escalated to product event due to reason for the return.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis:an overheating test could not be conducted due to another device malfunction.The likely cause was identified as bearings detached.It was also noted laser markings are partially illegible and both color rings are scratched and discolored.B3: date of event notification: 26th july 2024.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|