• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SERF BI MENTUM PFRK PLUS CUP 53

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SERF BI MENTUM PFRK PLUS CUP 53 Back to Search Results
Catalog Number DS5050153
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/21/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that patient was revised due to infection.Doi: unknown.Dor: unknown.Affected side: unknown.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.Although distributed by medtech orthopaedics joint reconstruction, the product is designed, manufactured, and labeled by the manufacturing supplier.The requirement of fda reporting, complaint investigation, root cause, and corrective action is the responsibility of the designing supplier of this item, per the supplier agreement.The product/information will be transferred to the supplier with the complaint problem statement for investigation.The results of the supplier investigation are to be populated into the medtech orthopaedics customer quality complaint management system.Per the supplier agreement, the supplier is responsible for any corrective actions identified during the complaint investigation process.An analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Event or Problem Description
Additional information received states that the affected side was the right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: although distributed by medtech orthopaedics joint reconstruction, the product is designed, manufactured, and labeled by the manufacturing supplier.The requirement of fda reporting, complaint investigation, root cause, and corrective action is the responsibility of the designing supplier of this item, per the supplier agreement.The product/information will be transferred to the supplier with the complaint problem statement for investigation.The results of the supplier investigation are to be populated into the medtech orthopaedics customer quality complaint management system.Per the supplier agreement, the supplier is responsible for any corrective actions identified during the complaint investigation process.The following investigation was provided by the supplier legal manufacturer: since the devices were not returned to the supplier, a technical investigation is not feasible.The investigation is therefore conducted using the available records for the affected batches and the monitoring records from the production period of the affected batches.An investigation focused on the steps that could impact the cleanliness and sterility of the products was conducted.The data recorded by the supplier throughout the manufacturing of the affected batches, the microbiological monitoring data, and the process monitoring data demonstrate that the products were clean and sterile when they were placed on the market.Therefore, the supplier devices cannot be the source of the infection that occurred.As part of medtech orthopaedics quality process all devices are manufactured, inspected, and released to approved specifications.No further investigation with regard to this complaint is required.The supplier will continue to monitor for trends.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: b5, d6a, d6b, d10 concomitant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BI MENTUM PFRK PLUS CUP 53
Common Device Name
BI MENTUM PFRK PLUS CUP 53
Manufacturer (Section D)
SERF
85 avenue des bruyères
décines-charpieu 69150
FR  69150
MDR Report Key20191572
Report Number1818910-2024-150022
Device Sequence Number7058871
Product Code MEH
UDI-Device Identifier03662200015153
UDI-Public(01)03662200015153(17)240331(10)1902177A
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section F) 08/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Catalogue NumberDS5050153
Device Lot Number1902177A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date06/02/2025
Event Location Hospital
Date Report to Manufacturer08/27/2024
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date09/11/2024
Supplement Report FDA Received Date10/17/2024
06/04/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/06/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BI MENTUM PFRK PE LINER 28 53.; BI MENTUM PFRK SCREW 5 65.; DELTA CER HEAD 12/14 28MM +1.5.; RECLAIM DISTAL TAPERED 20X140.; RECLAIM PRX BDY CONE 20X85.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexMale
Patient Weight78 KG
-
-