|
EXACTECH, INC. NV GXL LINR, NTRL, 36MM ID, GROUP 3 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
130-36-53 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Scar Tissue (2060); Osteolysis (2377); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); Bursitis (4832)
|
| Date of Event |
05/02/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: concomitant devices: ((b)(6)) 188-01-06 - wedge plasma x/o sz 6.((b)(6)) 180-01-56 - nv crown cup clstr hole 56mm group 3.((b)(6)) 170-36-00 - biolox delta femoral head 36mm od, +0mm.((b)(6)) 180-65-25 - alteon 6.5mm screw, 25mm.((b)(6)) 180-65-20 - alteon 6.5mm screw, 20mm.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 88 months after a left total hip replacement procedure, the patient underwent a revision procedure to address prosthesis wear.Revision surgery operative notes were provided.Patient left the operating room in stable condition.No further issues or complications were reported.No additional information is available.".
|
| |
|
Additional Manufacturer Narrative
|
|
Multiple mdr reports were filed for this event, please see associated reports: 1038671-2025-01315, 1038671-2025-01313.This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: h6 (type of investigation).The following sections were corrected: h6 (health effect - clinical code).The most likely underlying cause for the revision reported in is prosthesis wear and loosening and a combination of risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed from the reported information and the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|