| Model Number |
8888145040P |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during the procedure, there was a crack on the end piece or connecting piece of the reported product.There was a leak on the exit site.A hair crack at luer hubs was the unusual thing that was observed on the product.Flushing was done with no abnormalities.The catheter was not repaired.Tego was not utilized.Both sides of luer adapter had an issue.No excessive force is used on the device.No other defects or damages were found on the product.The product being utilized with the device was the dialysis blood tubing set.Octenisept was the cleaning agent used on the device.The insertion site was treated prior to product placement with octenisept.The remedial action performed was to repair with two repair kits.The procedure and treatment were able to proceed after the resolution was done.There was no blood loss, and blood transfusion was not required.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted two luer adapters that had been cut right before the extension tube.No abnormalities were spotted on the arterial luer adapter; however, a large crack was spotted on the side of the venous luer adapter.It was reported that the luer adapter was leaking and cracked.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|