| Model Number |
VTICMO 12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation not associated with a low vault.The lens was repositioned.In a separate visit the lens was explanted and the problem was resolved.It has been noted that the surgeon selected a different lens size than that initially suggested by the icl calculation software.Therefore, there is not enough safety and effectiveness data to support implantation in this patient category.
|
| |
|
Additional Manufacturer Narrative
|
|
H6- investigation type code: 4110- lens work order search-no similar complaint event(s) within associated lots were found.H11- manufacturer narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: h3 type of investigation code: 10 device evaluation: the lens was returned dry in a microcentrifuge vial with residue/ debris on product.Visual inspection found no damage.Dimensional inspection found the lens to be within specifications.(b)(4).
|
| |
|