SUNMED HOLDINGS LLC SUNMED; GREENLINE SUNBRITE LED PENLITE HANDLE Back to Search Results
Model Number 5-0236-90, 5-0236-99
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
All information reasonably known as of 11 sept 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an airlife product is defective or caused serious injury.
 
Event or Problem Description
Light flickers.We have been having problems with flickering in our greenline sunbrite led handles (5-0236-99 and 5-0236-90).Have you had any other complaints? one of the doctors said if he loosens the bulb it doesn't flicker.It's like there is too much distance between the battery and the bulb.Typical use,.
 
Additional Manufacturer Narrative
All information reasonably known as of 11 sept 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an airlife product is defective or caused serious injury.Per wi-20020-c1 rev 2, sec 15.6: for products in which airlife distributes, but is not the manufacturer, complaints will be captured and disseminated to the manufacturer but does not require investigation, reportability assessment, capa escalation, risk analysis, as this is the responsibility of the legal manufacturer and will be closed.
 
Event or Problem Description
Light flickers.We have been having problems with flickering in our greenline sunbrite led handles (5-0236-99 and 5-0236-90).Have you had any other complaints? one of the doctors said if he loosens the bulb it doesn't flicker.It's like there is too much distance between the battery and the bulb.Typical use,.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUNMED
Common Device Name
GREENLINE SUNBRITE LED PENLITE HANDLE
Manufacturer (Section D)
SUNMED HOLDINGS LLC
2710 northridge dr nw suite a
grand rapids MI 49544
Manufacturer (Section G)
SUNMED AP, LTD.
6-9 zhufeng rd., qiannan indus
qianwu town, doumen district
zhuhai guangdong, china 51917 0
CH   519170
Manufacturer Contact
melissa brickley
2710 northridge dr nw suite a
grand rapids, MI 49544
6162598415
MDR Report Key20195814
Report Number1314417-2024-00027
Device Sequence Number7583999
Product Code CCW
UDI-Device Identifier10889483008881
UDI-Public(01)10889483008881
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5-0236-90, 5-0236-99
Device Catalogue Number5-0236-90, 5-0236-99
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2024
Supplement Date Received by Manufacturer08/16/2024
Initial Report FDA Received Date09/11/2024
Supplement Report FDA Received Date09/23/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-