| Model Number |
DBEC-125 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Myocardial Infarction (1969)
|
| Date of Event |
05/22/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The diamondback 360 coronary orbital atherectomy device was used for treatment of 80% stenosed, severely calcified, 3mm-diameter lesion in proximal circumflex (cx) artery via right femoral approach.Three high-speed treatments were performed at the target lesion.Balloon angioplasty and stent placement were performed to complete the procedure.The clinical event committee reviewed procedural images and concluded that the diamondback 360 coronary orbital atherectomy device possibly contributed to a myocardial infarction.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 investigation conclusion code 22: the diamondback 360® coronary orbital atherectomy system instructions for use manual states that myocardial infarction is a potential adverse event that may occur and/or require intervention with use of the system.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.(b)(4).
|
| |
|
Search Alerts/Recalls
|