|
Sorin group italia has received a report that, during a procedure, two inspire oxygenators did not function as expected.Medical team elected to change out the first oxygenator (livanova ref (b)(4) ).With the second oxygenator, perfusionist elected to complete the case despite also the second oxygenator (present case, livanova ref (b)(4) ) was not functioning as expected.Livanova is submitting two different initial manufacturer reports for the two complained oxygenators.There is no report of any patient injury.
|
|
H11: based on the perfusion protocol received from the customer and analyzed, the maximum delta pressure value across the oxygenator membrane was 343 mmhg, measured with a pump flow rate of 6,27 lpm.In addition, analysis of the pump sheet revealed that the arterial flow was progressively reduced to mitigate the high pressure excursion.Through follow-up activity, no further information about any gas exchange issue possibly occurred or any additional medical intervention possibly conducted during the event could be retrieved.The date of the event was confirmed to be july 30th, 2024.Therefore, based on the information available and pump sheet analysis, the only issue occurred was an high pressure excursion.Present complaint has been then re-assessed as not reportable.Furthermore, the involved unit was returned to sorin group italy for inspection and investigation.After decontamination by gamma-rays (as per livanova procedure on blood contaminated goods), returned unit was found badly clogged when visually inspected, with large clots and dried blood residues along the oxygenator fibers.An extensive cleaning operation was carried out on the device by flushing the blood compartment with demineralized water repeatedly, but it proved ineffective to purge the fibers from the observed biological aggregates.In detail, during this preliminary step, a restricted flow rate was reproduced due to the high hydraulic resistance to the passage of water exhibited by the obstructed fibers.Based on the above findings, no simulated functional test with bovine blood could be possible.As per the conditions of the returned device, and taking into account the investigation results of previous similar cases, the reported event was traced back to the unexpected and progressive build up of biological material (platelet aggregates and fibrin mass) inside the oxygenator fibers that arose during the case at customer's facility.No manufacturing quality deviation possibly linked with the occurred phenomenon was established.Livanova conducted a literature and technical analysis about the trans-membrane oxygenator pressure gradients.Pressure drop excursion across the oxygenator is a known phenomenon reported in literature in all membrane oxygenators.The main cause of the pressure excursion has been identified in the platelet activation and adhesion within the oxygenator that progresses to increase resistance to blood flow during cardiopulmonary bypass.The factors influencing pressure excursions can be assigned to three main areas: - surgery clinical impact - cpb clinical impact - pathological patient conditions research on the phenomenon consistently concludes that the pressure excursion is complex and multi-factorial and in most of the cases none of the factors considered singularly causes the phenomenon.An accurate analysis and monitoring of the factors identified can lead to the reduction of the phenomenon during cpb.Nevertheless at current state of knowledge the complex nature of the phenomenon does not allow its complete elimination.
|