• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 44MM/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 44MM/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 6260-9-244
Medical Device Problem Codes Degraded (1153); Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Pain (1994); Ambulation or Postural Difficulties (2544); Metal Related Pathology (4530)
Date of Event 08/22/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Pt.Called and reported that they have a recalled hip and have to have it replaced.It's osteonics accolade system and it was done (b)(6) 2013.Insurance called and said that hip was recalled.I've really suffered.The pain is really bad and can hardly walk.Pending revision.
 
Event or Problem Description
Pt.Called and reported that they have a recalled hip and have to have it replaced.It's osteonics accolade system and it was done (b)(6) 2013.Insurance called and said that hip was recalled.I've really suffered.The pain is really really bad and can hardly walk.Pending revision.
 
Additional Manufacturer Narrative
Product code d2b common device name d2a corrected as per catalog number received.An event regarding pain involving a metal head was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: a review of medical records with a clinical consultant "conclusion/assessment: the patient had a left total hip arthroplasty performed (b)(6) 2013.No records of the patient's preoperative care or conservative management were presented.She has multiple comorbid conditions including schizophrenia narcotic use back pain with some radicular pain in a number of complicated medications.She has had prior total knee arthroplasties.She did apparently have a hip joint injection that relieved her pain for a very short time.She has noted to have arthritis on the opposite side as well.The total hip appears to been performed without complication multiple records from her hospital stay, most of which were noncontributory, do not show any complications while hospitalized.No radiographs pre or postoperatively, and no pre or postoperative medical records were provided for review.There is no documented complaints of the patient relative to the hip outside of the pi report.Event confirmation: a left total hip arthroplasty with a stryker implant can be confirmed from implant stickers.The implants: trident psl ha cluster shell 56¿f, 44 mm x3 0 degree polyethylene, accolade tmz f+ 127 degree v40 #4.5, lfit anatomic v40 femoral head 44 mm +4 mm can be confirmed.Recall of the implants cannot be confirmed.The patient complaints and the pi report cannot be confirmed.Root cause: the root cause for the total hip arthroplasty was hip arthritis.The root cause for the patient complaints cannot be ascertained as the complaints cannot be confirmed.A root cause for a recall cannot be ascertained as this cannot be confirmed.Of note the implants are labeled as osteonics rather than stryker by the patient." -product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.Although the reported lot code is in the scope of the capa, the reported failure mode is not in the scope of this recall.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
 
Event or Problem Description
Pt.Called and reported that they have a recalled hip and have to have it replaced.It's osteonics accolade system and it was done (b)(6) 2013.Insurance called and said that hip was recalled.I've really suffered.The pain is really bad and can hardly walk.Pending revision.Update: pt.Was revised on (b)(6) 2024.
 
Additional Manufacturer Narrative
Reported event: an event regarding pain involving a metal head was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of medical records with a clinical consultant "conclusion/assessment: the patient had a left total hip arthroplasty performed (b)(6) 2013.No records of the patient's preoperative care or conservative management were presented.She has multiple comorbid conditions including schizophrenia narcotic use back pain with some radicular pain in a number of complicated medications.She has had prior total knee arthroplasties.She did apparently have a hip joint injection that relieved her pain for a very short time.She has noted to have arthritis on the opposite side as well.The total hip appears to been performed without complication multiple records from her hospital stay, most of which were noncontributory, do not show any complications while hospitalized.No radiographs pre or postoperatively and no pre or postoperative medical records were provided for review.There is no documented complaints of the patient relative to the hip outside of the pi report.Event confirmation: a left total hip arthroplasty with a stryker implant can be confirmed from implant stickers.The implants: trident psl ha cluster shell 56¿f, 44 mm x3 0 degree polyethylene, accolade tmz f+ 127 degree v40 #4.5, lfit anatomic v40 femoral head 44 mm +4 mm can be confirmed.Recall of the implants cannot be confirmed.The patient complaints and the pi report cannot be confirmed.Root cause: the root cause for the total hip arthroplasty was hip arthritis.The root cause for the patient complaints cannot be ascertained as the complaints cannot be confirmed.A root cause for a recall cannot be ascertained as this cannot be confirmed.Of note the implants are labeled as osteonics rather than stryker by the patient." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Pt.Called and reported that they have a recalled hip and have to have it replaced.It's osteonics accolade system and it was done (b)(6) 2013.Insurance called and said that hip was recalled.I've really suffered.The pain is really bad and can hardly walk.Pending revision.Update: pt.Was revised on (b)(6)2024.
 
Additional Manufacturer Narrative
Reported event: an event regarding abnormal ion level involving a metal head was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: a review of medical records with a clinical consultant "conclusion/assessment: the patient had a left total hip arthroplasty performed (b)(6)2013.No records of the patient's preoperative care or conservative management were presented.She has multiple comorbid conditions including schizophrenia narcotic use back pain with some radicular pain in a number of complicated medications.She has had prior total knee arthroplasties.She did apparently have a hip joint injection that relieved her pain for a very short time.She has noted to have arthritis on the opposite side as well.The total hip appears to been performed without complication multiple records from her hospital stay, most of which were non-contributory, do not show any complications while hospitalized.No radiographs pre or postoperatively and no pre or postoperative medical records were provided for review.There is no documented complaints of the patient relative to the hip outside of the pi report.The med review addendum indicated [.] trunnion was found to "look fine" but there were black markings but no significant corrosion or structural damage.Frozen sections were "mostly normal showing a small number of neutrophils and some brown soft tissue fragments of indeterminate significance.Her follow-up on (b)(6)2024 showed well aligned implants hip that looked fine and hip was warm and dry.She was recommended to continue for therapy for 6 weeks.There are no radiographs provided for review.The final pathologic report was not provided for review.In short, the patient who had multiple other medical and social issues, most significantly for significant low back pain and stenosis, underwent revision hip surgery for increased metal levels and increased fluid collection around the hip on mri and based on presence of a large cobalt chrome lfit head which was on a tmzf stem.By report there had been a recall of the implant which was not otherwise documented.Event confirmation: a left total hip arthroplasty with a stryker implant can be confirmed.The implants: trident psl ha cluster shell 56¿f, 44 mm x3 0-degree polyethylene, accolade tmzf+ 127-degree v40 #4.5, lfit anatomic v40 femoral head 44 mm +4 mm can be confirmed.Recall of the implants cannot be confirmed.Increased metal levels can be confirmed.Revision surgery can be confirmed.Trunnionosis/corrosion cannot be confirmed.Root cause: the root cause for the total hip arthroplasty was hip arthritis.The root cause for the patient complaints cannot be ascertained as there were a number of comorbid conditions.A root cause for a recall cannot be ascertained as this was not confirmed.The root cause of the revision was patient pain, increased metal levels, assertion of a recall, and suspected trunnionosis.[.].Of note the implants are labeled as osteonics rather than stryker by the patient." event confirmation: a left total hip arthroplasty with a stryker implant can be confirmed from implant stickers.The implants: trident psl ha cluster shell 56¿f, 44 mm x3 0 degree polyethylene, accolade tmz f+ 127 degree v40 #4.5, lfit anatomic v40 femoral head 44 mm +4 mm can be confirmed.Recall of the implants cannot be confirmed.The patient complaints and the pi report cannot be confirmed.-product history review: review of the device history records indicate 5 devices were manufactured and accepted into final stock on 11/10/2010 with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: the med review addendum indicated [.] trunnion was found to "look fine" but there were black markings but no significant corrosion or structural damage.Frozen sections were "mostly normal showing a small number of neutrophils and some brown soft tissue fragments of indeterminate significance.Her follow-up on (b)(6)2024 showed well aligned implants hip that looked fine and hip was warm and dry.She was recommended to continue for therapy for 6 weeks.There are no radiographs provided for review.The final pathologic report was not provided for review.In short, the patient who had multiple other medical and social issues, most significantly for significant low back pain and stenosis, underwent revision hip surgery for increased metal levels and increased fluid collection around the hip on mri and based on presence of a large cobalt chrome lfit head which was on a tmzf stem.By report there had been a recall of the implant which was not otherwise documented.Event confirmation: a left total hip arthroplasty with a stryker implant can be confirmed.The implants: trident psl ha cluster shell 56¿f, 44 mm x3 0-degree polyethylene, accolade tmzf+ 127-degree v40 #4.5, lfit anatomic v40 femoral head 44 mm +4 mm can be confirmed.Recall of the implants cannot be confirmed.Increased metal levels can be confirmed.Revision surgery can be confirmed.Trunnionosis/corrosion cannot be confirmed.Root cause: the root cause for the total hip arthroplasty was hip arthritis.The root cause for the patient complaints cannot be ascertained as there were a number of comorbid conditions.A root cause for a recall cannot be ascertained as this was not confirmed.The root cause of the revision was patient pain, increased metal levels, assertion of a recall, and suspected trunnionosis.[.] no further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Pt.Called and reported that they have a recalled hip and have to have it replaced.It's osteonics accolade system and it was done (b)(6) 2013.Insurance called and said that hip was recalled.I've really suffered.The pain is really really bad and can hardly walk.Pending revision.Update: pt.Was revised on (b)(6) 2024.
 
Additional Manufacturer Narrative
Reported event: an event regarding abnormal ion level involving a metal head was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of medical records with a clinical consultant "conclusion/assessment: the patient had a left total hip arthroplasty performed (b)(6) 2013.No records of the patient's preoperative care or conservative management were presented.She has multiple comorbid conditions including schizophrenia narcotic use back pain with some radicular pain in a number of complicated medications.She has had prior total knee arthroplasties.She did apparently have a hip joint injection that relieved her pain for a very short time.She has noted to have arthritis on the opposite side as well.The total hip appears to been performed without complication multiple records from her hospital stay, most of which were noncontributory, do not show any complications while hospitalized.No radiographs pre or postoperatively, and no pre or postoperative medical records were provided for review.There is no documented complaints of the patient relative to the hip outside of the pi report.The med review addendum indicated [.] trunnion was found to "look fine" but there were black markings but no significant corrosion or structural damage.Frozen sections were "mostly normal showing a small number of neutrophils and some brown soft tissue fragments of indeterminate significance.Her follow-up on (b)(6) 2024 showed well aligned implants hip that looked fine and hip was warm and dry.She was recommended to continue for therapy for 6 weeks.There are no radiographs provided for review.The final pathologic report was not provided for review.In short, the patient who had multiple other medical and social issues, most significantly for significant low back pain and stenosis, underwent revision hip surgery for increased metal levels and increased fluid collection around the hip on mri and based on presence of a large cobalt chrome lfit head which was on a tmzf stem.By report there had been a recall of the implant which was not otherwise documented.Event confirmation: a left total hip arthroplasty with a stryker implant can be confirmed.The implants: trident psl ha cluster shell 56¿f, 44 mm x3 0-degree polyethylene, accolade tmzf+ 127-degree v40 #4.5, lfit anatomic v40 femoral head 44 mm +4 mm can be confirmed.Recall of the implants cannot be confirmed.Increased metal levels can be confirmed.Revision surgery can be confirmed.Trunnionosis/corrosion cannot be confirmed.Root cause: the root cause for the total hip arthroplasty was hip arthritis.The root cause for the patient complaints cannot be ascertained as there were a number of comorbid conditions.A root cause for a recall cannot be ascertained as this was not confirmed.The root cause of the revision was patient pain, increased metal levels, assertion of a recall, and suspected trunnionosis.[.] event confirmation: a left total hip arthroplasty with a stryker implant can be confirmed from implant stickers.The implants: trident psl ha cluster shell 56¿f, 44 mm x3 0 degree polyethylene, accolade tmz f+ 127 degree v40 #4.5, lfit anatomic v40 femoral head 44 mm +4 mm can be confirmed.Recall of the implants cannot be confirmed.The patient complaints and the pi report cannot be confirmed.Root cause: the root cause for the total hip arthroplasty was hip arthritis.The root cause for the patient complaints cannot be ascertained as the complaints cannot be confirmed.A root cause for a recall cannot be ascertained as this cannot be confirmed.Of note the implants are labeled as osteonics rather than stryker by the patient." updated addendum indicated [.] event confirmation: a left total hip arthroplasty with a stryker implant can be confirmed.The implants: trident psl ha cluster shell 56¿f, 44 mm x3 0-degree polyethylene, accolade tmzf+ 127-degree v40 #4.5, lfit anatomic v40 femoral head 44 mm +4 mm can be confirmed.Recall of the implants cannot be confirmed.Increased metal levels can be confirmed.Revision surgery can be confirmed.Trunnionosis/corrosion cannot be confirmed.Root cause: the root cause for the total hip arthroplasty was hip arthritis.The root cause for the patient complaints cannot be ascertained as there were a number of comorbid conditions.A root cause for a recall cannot be ascertained as this was not confirmed.The root cause of the revision was patient pain, increased metal levels, assertion of a recall, and suspected trunnionosis.[.] product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the med review addendum indicated [.] trunnion was found to "look fine" but there were black markings but no significant corrosion or structural damage.Frozen sections were "mostly normal showing a small number of neutrophils and some brown soft tissue fragments of indeterminate significance.Her follow-up on (b)(6) 2024 showed well aligned implants hip that looked fine and hip was warm and dry.She was recommended to continue for therapy for 6 weeks.There are no radiographs provided for review.The final pathologic report was not provided for review.In short, the patient who had multiple other medical and social issues, most significantly for significant low back pain and stenosis, underwent revision hip surgery for increased metal levels and increased fluid collection around the hip on mri and based on presence of a large cobalt chrome lfit head which was on a tmzf stem.By report there had been a recall of the implant which was not otherwise documented.Event confirmation: a left total hip arthroplasty with a stryker implant can be confirmed.The implants: trident psl ha cluster shell 56¿f, 44 mm x3 0-degree polyethylene, accolade tmzf+ 127-degree v40 #4.5, lfit anatomic v40 femoral head 44 mm +4 mm can be confirmed.Recall of the implants cannot be confirmed.Increased metal levels can be confirmed.Revision surgery can be confirmed.Trunnionosis/corrosion cannot be confirmed.Root cause: the root cause for the total hip arthroplasty was hip arthritis.The root cause for the patient complaints cannot be ascertained as there were a number of comorbid conditions.A root cause for a recall cannot be ascertained as this was not confirmed.The root cause of the revision was patient pain, increased metal levels, assertion of a recall, and suspected trunnionosis.[.] updated addendum indicated [.] event confirmation: a left total hip arthroplasty with a stryker implant can be confirmed.The implants: trident psl ha cluster shell 56¿f, 44 mm x3 0-degree polyethylene, accolade tmzf+ 127-degree v40 #4.5, lfit anatomic v40 femoral head 44 mm +4 mm can be confirmed.Recall of the implants cannot be confirmed.Increased metal levels can be confirmed.Revision surgery can be confirmed.Trunnionosis/corrosion cannot be confirmed.Root cause: the root cause for the total hip arthroplasty was hip arthritis.The root cause for the patient complaints cannot be ascertained as there were a number of comorbid conditions.A root cause for a recall cannot be ascertained as this was not confirmed.The root cause of the revision was patient pain, increased metal levels, assertion of a recall, and suspected trunnionosis.[.].No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 44MM/+4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key20199652
Report Number0002249697-2024-01259
Device Sequence Number6750201
Product Code JDI
UDI-Device Identifier07613327032758
UDI-Public07613327032758
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K061434
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 04/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2025
Device Catalogue Number6260-9-244
Device Lot NumberMJPJNN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/22/2024
Supplement Date Received by Manufacturer10/17/2024
11/09/2024
02/08/2025
03/25/2025
Initial Report FDA Received Date09/12/2024
Supplement Report FDA Received Date11/06/2024
12/04/2024
03/05/2025
04/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/10/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2249697-08/29/2016-007R
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention; Hospitalization;
Patient Age80 YR
Patient SexFemale
Patient Weight68 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-