| Catalog Number |
5343230 |
| Medical Device Problem Codes |
Restricted Flow rate (1248); Suction Problem (2170); Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiration date: 02/2025) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a dialysis catheter placement procedure via the right internal jugular vein, there was allegedly a kink about two centimeters from the cuff under fluoroscopy imaging.It was further reported that the flow rate was allegedly found to be slow to drain, and so the catheter was removed.Furthermore, there were allegedly some problems in the lumen on the suction side.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during a dialysis catheter placement procedure via the right internal jugular vein, there was allegedly a kink about two centimeters from the cuff under fluoroscopy imaging.It was further reported that the flow rate was allegedly found to be slow to drain, and so the catheter was removed.Furthermore, there were allegedly some problems in the lumen on the suction side.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported catheter kink, aspiration issue and restricted flow rates issues as no objective evidence was provided for review.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|