It was reported that during surgery, the zimmer blade was bowing creating a gap in the middle.When corrected, the gap was on either side of the blade.When skin was removed it is done so unevenly and in strips rather than in one piece.The patient required additional time under anesthesia and additional skin grafting was required to complete the surgery.The harvest site did not require sutures.Another product was not used to complete the surgery.Due diligence is complete, there is no additional information available.
|
|
This event has been recorded by zimmer biomet under (b)(4).The following sections have been corrected/updated: b4, b5, d2, d4, g1, g3, g6, h1, h2, h3, h4, h6, h10, and h11.No product was returned or pictures provided; functional, visual, and/or dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|