|
Updated fields: h3, h6 and h11.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation foreign material in the device could not be identified, there was no physical damage where the foreign material was found, and there were no reported reprocessing deviations from the ifu.It is possible the foreign material was not removed during reprocessing.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu) section below: ifu states the detection method in gastrointestinal videoscope, gif-lv1, colonovideoscope, cf-lv1l, cf-lv1i, operation manual chapter 3 preparation and inspection.Ifu states the preventive measure in gastrointestinal videoscope, gif-lv1, colonovideoscope, cf-lv1l, cf-lv1i, reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories).Olympus will continue to monitor field performance for this device.
|