| Model Number |
SWVLK TENO BIO-COMP 7X 19.5MM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a biceps tenodesis surgery the anchor broke off.There was no harm for patient, operator or third party.The surgery was finished successfully with a different device, ar-1662bc-8.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to improper bone prep; misaligned insertion; prying/leveraging the device during insertion.
|
| |
|
Event or Problem Description
|
|
A piece did break off inside the patient, but it was retrieved successfully.
|
| |
|
Search Alerts/Recalls
|