MENTOR TEXAS MENTOR SILTEX ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3542635 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
08/19/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation revision surgery with implantation of a 250cc mentor siltex round moderate profile saline breast implant prosthesis.Post-operatively, the patient was diagnosed with a deflated right breast implant during an in-office visit with a medical professional.As a result, the patient underwent unilateral right-sided removal and replacement with an undisclosed breast implant on (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.D4: udi: (01)gtin unavailable, product made prior to gtin compliance date.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: deflation manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On october 09, 2024, the mentor analysis lab received the suspect medical device for evaluation.On october 11, 2024, mentor completed an evaluation on the returned device.Mentor conducted visual inspection of the device.Visual analysis of the returned sample revealed that the breast implant had a tear within an area of siltex cracking on the posterior view, measuring approximately 0.3 cm.The evaluation determined that the possible cause of the deflation is consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.On october 14, 2024, mentor became aware information was omitted from the initial report.Corrected data: the gtin should have been reported in the initial report.D4 primary udi number: (b)(4).D4: udi: product made prior to udi compliance date.Udi not required.Manufacturer¿s reference number: (b)(4).
|
| |
|