Model Number 3256
Medical Device Problem Codes Unstable (1667); Mechanics Altered (2984); Intermittent Loss of Power (4016)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that unstable speed occurred.The 90% stenosed target lesion was located in the mildly tortuous and severely calcified circumflex artery.A 1.50mm, 1.25mm rotapro and console were selected for use.During preparation, the 1.50mm burr stalled briefly.When it went into the patient with a set speed of 165k.The speed reached at a maximum of 200k, and it stalled multiple times and would no longer run as it stuck in the proximal lesion.The device was removed from the body and was changed to a 1.25 burr.When placed prior to the lesion at a set speed of 165k.The rotation speed per minute (rpm) appeared on the screen with a maximum 0f 220k, but no sound of the burr was heard, and no stalled signal was on.The burr was also got stuck in the lesion.The device was removed and completed the procedure with a different device.The procedure got prolonged, but no patient complications were reported.
 
Additional Manufacturer Narrative
Investigation results: device analysis findings: the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record (dhr) review: it was confirmed that this device met manufacturing specification prior to distribution and there no manufacturing deviations which could have contributed to the reported event.Risk review: a risk review was completed and confirmed that the event of console- erratic speed (normal mode rotation - speed unstable) and system stall (system - speed too slow) were defined in the risk documentation.These event type have been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion: based on a thorough review of the reported complaint, no definitive product-related cause could be identified.All available documentation and evidence were evaluated; however, the investigation did not confirm any specific failure mode, contributing factor, or condition that could explain the reported allegation.Therefore, the conclusion code assigned to this investigation is "cause not established".
 
Event or Problem Description
It was reported that unstable speed occurred.The 90% stenosed target lesion was located in the mildly tortuous and severely calcified circumflex artery.A 1.50mm, 1.25mm rotapro and console were selected for use.During preparation, the 1.50mm burr stalled briefly.When it went into the patient with a set speed of 165k.The speed reached at a maximum of 200k, and it stalled multiple times and would no longer run as it stuck in the proximal lesion.The device was removed from the body and was changed to a 1.25 burr.When placed prior to the lesion at a set speed of 165k.The rotation speed per minute (rpm) appeared on the screen with a maximum 0f 220k, but no sound of the burr was heard, and no stalled signal was on.The burr was also got stuck in the lesion.The device was removed and completed the procedure with a different device.The procedure got prolonged, but no patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAPRO
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
150 baytech dr
san jose CA
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key20206075
Report Number2124215-2024-55894
Device Sequence Number7650084
Product Code MCX
UDI-Device Identifier08714729975076
UDI-Public08714729975076
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P900056/S166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3256
Device Catalogue Number3256
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2024
Supplement Date Received by Manufacturer03/30/2026
Initial Report FDA Received Date09/12/2024
Supplement Report FDA Received Date04/20/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-