On (b)(6) 2024, this patient underwent an endovascular treatment for infrarenal aortic pseudoaneurysm (just distal the right renal artery) using an aortic extender component of gore® excluder® conformable aaa endoprosthesis with active control system.After implanting a bare-metal stent in the left renal artery (chimney technique), (b)(6) was slowly deployed just distal the right renal artery; however, it moved proximally and the right renal artery was unintentionally covered for approximately 90%.An additional bare-metal stent was placed in the right renal artery.Also, an additional gore® viabahn® endoprosthesis was deployed to extend the left chimney.Triple kissing ballooning was performed, and no endoleak was found by the final angiography.The patient tolerated the procedure.
|
|
H3: code "other" was selected as the medical device remains implanted.Return not possible.H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.H6: code d12: according to the gore® excluder® conformable aaa endoprosthesis instructions for use (ifu), adverse events that may occur and/or require intervention include, but are not limited to: occlusion / stenosis of device or native vessel.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|