ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065000138
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is:(b)(4).
 
Event or Problem Description
A non healthcare professional reported that the phacoemulsification needle broke off during cataract surgery.The procedure was completed by replacing the phacoemulsification tip with new one.There was no patient harm.
 
Additional Manufacturer Narrative
Additional information provided in d.9, h.3, h.6, and h.11.One opened phacoemulsification tip and tip wrench in a small parts tray (smpt) was received for the report.Sample was visually inspected and found to be nonconforming.Phacoemulsification tip was observed to be broken at aspiration bypass (ab) hole; breakage is very jagged.Wall thickness is even.Wear observed on threads, back of flange, and nut corners consistent with threading on handpiece.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The photos attached to the file were reviewed by the manufacturing site.Photo show a broken phacoemulsification tip and a wrench.Another photo shows a label with reported lot number.Reported issue confirmed from photo.The complaint evaluation confirms the phacoemulsification tip was broken.The root cause for the broken phacoemulsification tip cannot be determined from this evaluation.The phacoemulsification tip visual inspections do not show any manufacturing issues that would cause the broken phacoemulsification tip.The exact root cause for the broken phacoemulsification tip is unknown.Therefore, specific action with regards to this complaint cannot be taken.All phacoemulsification tips are 100% visually inspected by trained operators using thirty times magnification during the manufacturing process.Any nonconformance, such as the broken phacoemulsification tip exhibited on the returned opened sample, are removed from the lot, and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20209279
Report Number1644019-2024-02083
Device Sequence Number7801370
Product Code LRO
UDI-Device Identifier00380650001382
UDI-Public(01)00380650001382(17)260430(11)240530(10)16U472
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Catalogue Number8065000138
Device Lot Number16U472
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/16/2024
Supplement Date Received by Manufacturer11/17/2024
Initial Report FDA Received Date09/13/2024
Supplement Report FDA Received Date12/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/30/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-