• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL; INTRAOCULAR LENS Back to Search Results
Model Number SY60WF
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is:(b)(4).
 
Event or Problem Description
A non health care professional reported that during surgery, they have noticed that the lens was cracked after implanting into the patient eye.Subsequently the lens was removed during initial procedure.Additional information have been received it states that the lens was removed and replaced with unspecified another lens with same type and diopter.There was no patient harm.
 
Additional Manufacturer Narrative
Additional information has been provided in h.3., h.6., and h.11 the product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The account indicated the use of qualified associated products.The product investigation could not identify a root cause for the reported complaint of cracked lens.The product has not been received to evaluate.The ifu (instructions for use) instructs to completely fill the cartridge with ovd (viscosurgical device) immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol (intraocular lens) to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.No further information has been provided at this time.File will be reopened when new information or the product is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20210125
Report Number1119421-2024-01806
Device Sequence Number7202512
Product Code HQL
UDI-Device Identifier00380652251587
UDI-Public(01)00380652251587(17)271231(11)240102(10)15775427
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSY60WF
Device Catalogue NumberSY60WF.250
Device Lot Number15775427
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/15/2024
Supplement Date Received by Manufacturer10/10/2024
Initial Report FDA Received Date09/13/2024
Supplement Report FDA Received Date11/05/2024
Date Device Manufactured01/02/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MCPHEARSON FORCEPS; MONARCH III IOL CARTRIDGE D; MONARCH III IOL INJECTOR; PROVISC
Patient Age67 YR
Patient SexFemale
-
-