| Model Number |
161005 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Hamilton medical ag event reference number: cer (b)(4).The internal product investigation is still ongoing.As soon as the results are available, the manufacturer will submit an update of this report unsolicited.
|
| |
|
Event or Problem Description
|
|
The event description according to the customer is as follows: customer complained of replacing expiratory valve assm and checkvalve more than normal every 3 months typical.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).It was reported that the customer complained of replacing expiratory valve assembly and check valve more than normal every 3 months typical.No log files were provided.Expiratory valve assembly and check valve assembly were replaced, and the customer were trained on utilizing proper humidifier shutdown, filtering and cleaning practices and the device was released back into use.
|
| |
|