| Catalog Number |
81324 |
| Medical Device Problem Codes |
Restricted Flow rate (1248); No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
While using a cavitron select sps g124 they allege that they have no water and the handpiece and handpiece cable are getting hot, no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
2 (two) unsuccessful requests to retrieve suspect product or investigation result have been made and documented.Complaint will be reopened if suspect product or investigation result arrives per 8000-sop-038.
|
| |
|
Search Alerts/Recalls
|