• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. NEXGEN ARTICULAR SURFACE SIZE EF 10 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. NEXGEN ARTICULAR SURFACE SIZE EF 10 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Synovitis (2094); Osteolysis (2377); Joint effusion (4849)
Date of Event 08/30/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: 00598603702 - stemmed nonaugmentable tibial component option cr/ps/lps size 4 for cemented use only - j6818739.00576401552 - femoral component option for cemented use only size e right - 63892742.00597206529 - all poly petella standard cemented size 29 mm diameter 8.0 mm thickness - 64774869.Unknown - unknown bone cement - unknown.Customer has indicated that the product will not be returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient underwent an initial right total knee arthroplasty.Approximately 2 years post-op, the patient was revised due to pain, swelling, effusion, and tibial loosening.During the revision, the surgeon noted synovitis, osteolysis and that the tibial component was grossly loose.The tibia, femur, and articular surface were exchanged without complications.The patella remains implanted.Attempts have been made and all available information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: x-ray showed early aseptic loosening right knee secondary to polyethylene debris, developed pain on medial aspect of right knee and doing fine now; office note exam was positive for swelling and effusion; op note stated the tibia was grossly loose at the cement bone interface, excised tibia component and noted osteolysis of medial tibia plateau below the cement, femur well fixed, excised due to protocol of osteolytic process.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXGEN ARTICULAR SURFACE SIZE EF 10 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key20213284
Report Number0001822565-2024-02946
Device Sequence Number7745631
Product Code JWH
UDI-Device Identifier00889024634923
UDI-Public(01)00889024634923(17)290303(10)65038733
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K173057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00596403210
Device Lot Number65038733
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2024
Supplement Date Received by Manufacturer01/24/2025
Initial Report FDA Received Date09/13/2024
Supplement Report FDA Received Date01/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-