• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT BROACH SZ4

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT BROACH SZ4 Back to Search Results
Catalog Number 257000100
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/03/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that upon inspection the following parts were found.Summit broach sz4 - not attaching to broach handle.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received: was any alleged device deficiency reported for the (620510101) impactor handle? the impactor handle had bad or damaged threads.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, ¿actis broach sz 6 - not attaching to broach handle, summit broach sz4 - not attaching to broach handle, actis broach sz8 - not attaching to broach handle, replacement suction cup - ripped, emphasys lnr imp tip 32 - cracked, s/c & mod cath trl 48/28 - faded¿.The product returned to medtech orthopaedics for evaluation.The medtech orthopaedics team conducted a visual inspection of the returned device.Visual analysis revealed that summit broach sz4 has the post slightly bent.Additionally, the overall surface of the post exhibits deep scratches and signs of heavy usage.No other defects that could impact the assembly of the broach were observed.Damage observed on the post is consistent with a component failure that was caused by exposure to unintended forces like impacting the device before it was fully seated with the mating handle or by use of excessive force in a prying motion and overloading the material.A functional test performed with a mating device laboratory sample revealed that the broach was not able to be fully assemble; most likely due to the observed condition of the post.The reported complain condition "unable to assemble" was able to be replicated.A dimensional inspection was not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the summit broach sz4 would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended use error, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code (a0805) provided is not a valid finished goods lot number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT BROACH SZ4
Common Device Name
BROACH
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20215181
Report Number1818910-2024-19301
Device Sequence Number12635984
Product Code HTQ
UDI-Device Identifier10603295142911
UDI-Public(01)10603295142911(10)A0805
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number257000100
Device Lot NumberA0805
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/03/2024
Supplement Date Received by Manufacturer09/20/2024
10/30/2024
Initial Report FDA Received Date09/13/2024
Supplement Report FDA Received Date09/25/2024
11/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/15/2005
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-