The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
H11: additional manufacturer narrative: updated: b4, d4, g3, g6, h2, h4, h6.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Calcific degeneration involves the gradual accumulation of calcium deposits on the valve leaflets.It is a disease continuum from sclerosis to chronic inflammation that leads to leaflet calcification.These calcium deposits, over time, cause the leaflets to become rigid and less flexible, which can account for failure modes such as stenosis and regurgitation.Svd is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv implantation.The common endpoint of all these factors is calcification and degradation.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration.An engineering evaluation is not required because there is no allegation of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no use-related issue with a hazardous situation; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regard to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.A definitive root cause cannot be conclusively determined; however, patient factors likely caused or contributed.
|