| Model Number |
4593 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438); Pacing Problem (1439); Under-Sensing (1661); Migration (4003)
|
| Health Effect - Clinical Codes |
Syncope/Fainting (4411); Twiddlers Syndrome (4563)
|
| Date of Event |
08/28/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited oversensing due to noise with high within range pacing impedance measurements.In addition, one inappropriate burst of anti-tachycardia pacing (atp) was delivered.The patient experienced a large amount of weight loss, as a result this cardiac resynchronization therapy defibrillator (crt-d) migrated and was unstable in pocket.However, the patient had manipulated the device causing noise recorded.The rv lead exhibited variable high within range pacing impedance measurements.Isometric tests were performed, and measurements were in range.The patient declined the device relocation and remains under monitoring.No adverse patient effects were reported.This rv lead remains in service.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited oversensing due to noise with high within range pacing impedance measurements.In addition, one inappropriate burst of anti-tachycardia pacing (atp) was delivered.The patient experienced a large amount of weight loss, as a result this cardiac resynchronization therapy defibrillator (crt-d) migrated and was unstable in pocket.However, the patient had manipulated the device causing noise recorded.The rv lead exhibited variable high within range pacing impedance measurements.Isometric tests were performed, and measurements were in range.The patient declined the device relocation and remains under monitoring.No adverse patient effects were reported.This rv lead remains in service.It was reported that this rv lead exhibited oversensing due to noise with pacing inhibition, resulting in inappropriate anti-tachycardia pacing (atp) and syncope.Rv lead pacing impedance was also fluctuating by 900 ohms every few days with out-of-range measurements observed.Undersensing was also noted.The patient was admitted to the hospital for a cardiology consult and to potentially replace the rv lead.The rv lead remains in service.No additional adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited oversensing due to noise with high within range pacing impedance measurements.In addition, one inappropriate burst of anti-tachycardia pacing (atp) was delivered.The patient experienced a large amount of weight loss, as a result this cardiac resynchronization therapy defibrillator (crt-d) migrated and was unstable in pocket.However, the patient had manipulated the device causing noise recorded.The rv lead exhibited variable high within range pacing impedance measurements.Isometric tests were performed, and measurements were in range.The patient declined the device relocation and remains under monitoring.No adverse patient effects were reported.This rv lead remains in service.It was reported that this rv lead exhibited oversensing due to noise with pacing inhibition, resulting in inappropriate anti-tachycardia pacing (atp) and syncope.Rv lead pacing impedance was also fluctuating by 900 ohms every few days with out-of-range measurements observed.Undersensing was also noted.The patient was admitted to the hospital for a cardiology consult and to potentially replace the rv lead.The rv lead remains in service.No additional adverse patient effects were reported.Additional information was received, this rv lead was tested at office and there were no signs of lead fracture.This rv lead remains in service.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: h6: absence of treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to update section b5 and h6 codes.Prv.Correction: h6: absence of treatment.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited oversensing due to noise with high within range pacing impedance measurements.In addition, one inappropriate burst of anti-tachycardia pacing (atp) was delivered.The patient experienced a large amount of weight loss, as a result this cardiac resynchronization therapy defibrillator (crt-d) migrated and was unstable in pocket.However, the patient had manipulated the device causing noise recorded.The rv lead exhibited variable high within range pacing impedance measurements.Isometric tests were performed, and measurements were in range.The patient declined the device relocation and remains under monitoring.No adverse patient effects were reported.This rv lead remains in service.It was reported that this rv lead exhibited oversensing due to noise with pacing inhibition, resulting in inappropriate anti-tachycardia pacing (atp) and syncope.Rv lead pacing impedance was also fluctuating by 900 ohms every few days with out-of-range measurements observed.Undersensing was also noted.The patient was admitted to the hospital for a cardiology consult and to potentially replace the rv lead.The rv lead remains in service.No additional adverse patient effects were reported.Additional information was received, this rv lead was tested at office and there were no signs of lead fracture.This rv lead remains in service.No additional adverse patient effects were reported.Additional information indicated that this rv lead was surgically abandoned.No additional adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited oversensing due to noise with high within range pacing impedance measurements.In addition, one inappropriate burst of anti-tachycardia pacing (atp) was delivered.The patient experienced a large amount of weight loss, as a result this cardiac resynchronization therapy defibrillator (crt-d) migrated and was unstable in pocket.However, the patient had manipulated the device causing noise recorded.The rv lead exhibited variable high within range pacing impedance measurements.Isometric tests were performed, and measurements were in range.The patient declined the device relocation and remains under monitoring.No adverse patient effects were reported.This rv lead remains in service.It was reported that this rv lead exhibited oversensing due to noise with pacing inhibition, resulting in inappropriate anti-tachycardia pacing (atp) and syncope.Rv lead pacing impedance was also fluctuating by 900 ohms every few days with out-of-range measurements observed.Undersensing was also noted.The patient was admitted to the hospital for a cardiology consult and to potentially replace the rv lead.The rv lead remains in service.No additional adverse patient effects were reported.Additional information was received, this rv lead was tested at office and there were no signs of lead fracture.This rv lead remains in service.No additional adverse patient effects were reported.Additional information indicated that this rv lead was surgically abandoned.No additional adverse patient effects were reported.The complaint device was not returned to bsc, therefore product analysis could not be performed.The primary reported observation is addressed in product labeling (e.G.Instructions for use).Therefore, well-understood factors likely contributed to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to update section b5 and h6 codes.Prv
correction:
h6: absence of treatment
this product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.Review of labeling determined that the complaint situation was listed in the manual.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of failure to follow instruction.
|
| |
|
Search Alerts/Recalls
|