H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
It was learned through implant patient registry that a 27mm 8300ab aortic valve in aortic position was explanted and replaced with a 27mm 11500a aortic valve after an implant duration of 6 years, 8 months due to unknown reasons.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional "customer complaint".The information reported may or may not be related to the edwards device.
|
|
It was learned through implant patient registry and pathology report that a 27mm 8300ab aortic valve in aortic position was explanted and replaced with a 27mm 11500a aortic valve after an implant duration of 6 years, 8 months due to degeneration and calcification.The patient was in recovery post-operative.Per pathology, gross examination showed leaflets to be calcified and stiff.One leaflet showed a 0.5cm penetrating defect near the suture ring.No vegetations.Microscopic examination showed valve tissue thickening, fibrinoid degeneration, and calcification.
|
|
H11: additional manufacturer narrative: updated sections: h6 (device code, type of investigation, investigation findings, and investigation conclusions).The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Engineering evaluation summary: per technical summary (b)(4), rev b, calcific degeneration involves the gradual accumulation of calcium deposits on the valve leaflets.It is a disease continuum from sclerosis to chronic inflammation that leads to leaflet calcification.These calcium deposits, over time, cause the leaflets to become rigid and less flexible, which can account for failure modes such as stenosis and regurgitation.Svd is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv implantation.The common endpoint of all these factors is calcification and degradation.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration.Per (b)(4), rev p, an engineering evaluation is not required because there is no allegation of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no use-related issue with a hazardous situation; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regard to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.Per (b)(4), rev p, and (b)(4), rev o, a capa/scar/pra is not required, as there are no confirmed product or labeling non-conformance's and no other triggers are met.A definitive root cause cannot be conclusively determined; however, patient factors likely caused or contributed.All pertinent information available to edwards lifesciences has been submitted.
|