• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. REMSTAR AUTOA-FLEX; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. REMSTAR AUTOA-FLEX; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number IN561HS
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2024
Type of Reportable Event Malfunction
Event or Problem Description
The device was returned to a third-party service center for evaluation in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap/ bipap, and mechanical ventilator devices.The device was not in patient use.There was no report of serious patient harm or injury.The internal part of the device was inspected visually and found evidence of visible foam particles inside the assembly.Secondary findings include dust/dirt contamination inside the device.In addition, the devices error log was downloaded, and one error code was present when reviewed.Lastly, the device has been scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTAR AUTOA-FLEX
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key20218415
Report Number2518422-2024-58818
Device Sequence Number7450874
Product Code BZD
UDI-Device Identifier00606959030763
UDI-Public(01)00606959030763
Combination Product (Y/N)N
Initial Reporter CountryMT
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 09/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberIN561HS
Device Catalogue NumberIN561HS
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2024
Initial Report FDA Received Date09/13/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/22/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-