| Model Number |
10CN10R |
| Medical Device Problem Codes |
Use of Device Problem (1670); Gas/Air Leak (2946)
|
| Health Effect - Clinical Codes |
Discomfort (2330); Unintended Extubation (4564)
|
| Date of Event |
07/19/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
According to the reporter, the balloon of the tracheostomy tube was deflated prematurely.The patient had to change the trach tube every week or every 15 days instead of once a month resulted to discomfort.Upon visual examination, the nurse reported that the balloon was 'wrinkled' and when it was inflated with syringe, a punctured balloon was observed.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the cuff of the tracheostomy tube was deflated prematurely.The patient had to change the trach tube every week or every 15 days instead of once a month resulted to discomfort.Upon visual examination, the nurse reported that the cuff was 'wrinkled' and when it was inflated with syringe, a punctured cuff was observed.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that there was a straight-line slit measuring 0.48 mm in length on the main body of the cuff.Functionally, an attempt made to inflate the returned unit failed as the cuff deflated.It was reported that the cuff of the tracheostomy tube was deflated prematurely.The patient had to change the trach tube every week or every 15 days instead of once a month resulted to discomfort.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: to ease insertion and to guard against cuff perforation from sharp edges of cartilage, taper back the cuff by deflating the cuff and gently moving it away from the distal tip of the cannula toward the neck plate as the residual air is removed by deflation.When tapering the cuff, do not use sharp instruments that would damage the cuff, such as forceps or hemostats.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|