• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 2.5 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 2.5 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-35-2509
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Osteolysis (2377)
Date of Event 09/10/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10 concomitant devices: (b)(6) 02-010-01-0225 - logic femoral ps cem left sz 2.5.(b)(6) 02-012-45-2525 - lgc tibial fit tray cem sz 2.5f / 2.5t.(b)(6) 200-02-35 - three peg patella 35mm.(b)(6) 203-96-43 - 11-4137 sryker sys 6 90x13x1.19.The product associated with the reported event is within the scope of recall z-0021-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that approximately 83 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, left knee pain; osteolysis, loosening of femur and tibia, polyethylene wear on the tibial insert.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, tibial loosening, and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.H6: corrected component and investigation clinical codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA PS MOD INSRT SZ 2.5 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville 32653
3523771140
MDR Report Key20223452
Report Number1038671-2024-03523
Device Sequence Number15404084
Product Code JWH
UDI-Device Identifier10885862001764
UDI-Public10885862001764
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K093360
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/12/2020
Device Catalogue Number02-012-35-2509
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2024
Supplement Date Received by Manufacturer05/21/2025
Initial Report FDA Received Date09/16/2024
Supplement Report FDA Received Date05/21/2025
Date Device Manufactured07/12/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-