• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD VERSASTEP PLUS; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD VERSASTEP PLUS; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number VS101011P
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a laparoscopic sleeve gastrectomy, at the beginning of the procedure, the rubber piece came off from trocar top.No device component fell into the cavity of the patient.There was no patient injury.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted only the flip-top converter and seal hosing were received.The seal for the flip-top converter was disengaged.The circular seal and seal housing were intact.It was reported that the seal disengaged.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when contact is made with a surgical device during clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: use special care when introducing or removing sharp-edged or sharp-angled endoscopic instruments to minimize the potential of inadve rtent damage to the seal.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSASTEP PLUS
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key20224127
Report Number9612501-2024-02324
Device Sequence Number7450076
Product Code GCJ
UDI-Device Identifier10884521082878
UDI-Public10884521082878
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K012539
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVS101011P
Device Catalogue NumberVS101011P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2024
Supplement Date Received by Manufacturer10/30/2024
Initial Report FDA Received Date09/16/2024
Supplement Report FDA Received Date11/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-