• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME Back to Search Results
Model Number KD-V411M-0320
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported the subject device's knife wire was broken.The issue occurred during preparation for use for a endoscopic sphincterotomy procedure that was completed using a similar device.There were no reports of patient harm.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and the device malfunction was confirmed.It is most likely the mechanism causing the cutting wire breakage might be the following: 1.The cutting wire where the wire coating was torn came into contact with the distal end of the endoscope when the forceps elevator was raised.2.Under circumstances described in note 1, an electric conduction was activated.This caused the cutting wire to become hot instantly at the contact point.As a result, the cutting wire broke.However, the definitive root cause of the reported issue could not be determined.The event can be detected/prevented by following the instructions for use (ifu): since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.When inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.Do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.If you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion.Be sure that the rear end of the cutting wire is extended from the distal end of the endoscope.Insufficient output or unintended tissue injury may occur in case the cutting wire contacts the forceps elevator.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN SPHINCTEROTOME V
Common Device Name
SINGLE USE 2-LUMEN SPHINCTEROTOME
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20224157
Report Number9614641-2024-01831
Device Sequence Number18015836
Product Code KNS
UDI-Device Identifier04953170382604
UDI-Public04953170382604
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K950166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-V411M-0320
Device Lot Number41V
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2024
Supplement Date Received by Manufacturer10/02/2024
Initial Report FDA Received Date09/16/2024
Supplement Report FDA Received Date10/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-