Visual inspection: during the investigation, scratches on the outer housing of the valve, but no significant deformations or damage was determined.Permeability test: a permeability test has shown that all components are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in vertical/ or both the horizontal as well as the vertical positions.The results show that the m.Blue operates within the accepted tolerance in both positions.Adjustment test: the m.Blue valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection: after dismantling of the valve, deposits were found in the m.Blue results: based on our investigation results, we can determine deposits.The deposits had no affect to the technical properties at the time of the investigation.The cause of the aforementioned functional impairment is not known to us at the time of the examination.Organic material in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspections when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|